T6000

WVTR Test System


T6000 system is an equipment for calcium test inspection. This system configures of the electrical test unit, the environment chamber, andoptionally the optical test unit. For multi channel system, it has the switching unit. It estimates the performance of OLED encapsulation, especially WVTR.

Evaluation of Display / Solar Encapsulation

Calcium Sample Degradation Test with Temperature and Humidity

Electrical Resistance / Conductance Measurement

Water Vapor Transmission Rate (WVTR) Calculation

Temperature / Humidity Chamber


System Specification

Description

Description

Specification

Test Sample

Standard

OLED Unit Cell

Channel

Standard

16CH

IV Measurement Unit

Range

Resolution

Test Current

Accuracy

Connection

Interface

100.0000 Ohm ~ 100.0000 MOhm

Min. 100uOhm

Max. 100mA

0.1%

4 Probe

RS-232

Switching Unit

Type

Capacity

MultiplexSwitch

1 to 16

Chamber(60L)

Temperature

Range

Fluctuation

Heating Time

Cooling Time

Internal Size : 660mm(W) x 1040mm(D) x 1065mm(H)

-20°C ~ 150°C (without Humidity)

±0.2°C @ 25°C / 60% RH

70min. (-20°C -> 120°C)

150min. (120°C -> -20°C)

Humidity

Type

Range

Fluctuation

Direct Water System

20%RH ~ 95%RH

±2.7%RH

Sample Mounting Unit

Material

Connection

Teflon or Aluminium

Pin Contact & Waterproof Connector

System Specification

System Configuration

  • Leaflet
Leaflet

T6000 WVTR Test System

Leaflet Download (0MB)

System Components

Multi-Channel Resistance Meter

Switching Unit

Chamber

Signal Amplifier

Signal Amplifier

T760 WVTR Test Software