Q1100
Cyclic Voltammetry Test System
The Cyclic Voltammetry is a basic experiment for the characterization of electrochemical systems. The long-cycle repeated measurement of Cyclic Voltammetry is widely used for quick evaluation of electrochemical materials for the application of devices such as rechargeable batteries, electrochemical double-layer capacitors and photo-electrochemical systems. Q1100 provides multi-channel management capability for efficient measurements of multiple sample devices.
Multi-Channel Cylic Voltammetry
High-Speed Data Acquisition
Sequence Builder
Individual Channel Management
Gas Control & Light Source Options
System Specification
Classification
Content
Channel #
4CH (Independent Configuration)
Test Board #
1EA
Mode
Voltage Control, Current Measurement
Output Connection
2Probe, 3Probe, 4Probe
Voltage Control
Range
Capability
Resolution
Accuracy
Stability
Scan Rate
Frequency
Setting Parameter
-10V ~ +10V
10mA
16bit DAC
< 0.1% of full scale
<0.1%/h after 30 min. warming up
10mV/sec ~ 100V/sec
0.001 ~ 100Hz
Voltage Sweep Range, Volatge Step, Scan Rate, Cycle
Current Measurement
Configuration
Range
Resolution
Accuracy
Stability
Sampling Rate
Measurement Parameter
Zero Resistance Ammeter
4 Decades (10μA, 100μA, 1mA, 10mA)
16bit ADC
< 0.01% of full scale
<0.1%/h after 30 min. warming up
40ksps
Cycle, Time, Voltage, Current, Epc / Ipc / Epa / Ipa
Time
1sec ~ 65,000 hrs
Measured Data
Voltage, Current, Time
Operating Temperature
20℃ ~ 30℃
PC Interface
Serial / LAN
Dimension
450mm(W) x 300mm(D) x 150mm(H)
Weight
< 10 Kg
Line Voltage
220V 1Ø 50 ~ 60Hz, <3A
System Specification
System Configuration
System Components
System Frame
Dark Box
Measurement Unit
Optional Gas Controller
Cell Kit
Manipulate Unit
Q730 Battery Parameter Test Software