M7000PLUS Auto I-V-L Test System is a system testing multi display devices with McScience Self-Developed Optical Detector (LUMO Imaging Spectroradiometer) and Source Measure Unit (Metro M131 LAB). By adjusting 3-axis position automatically, it helps to measure many devices as possible in a short experiment time and various test items cover wide range of display testing standards not only the basic display device characteristics, but also their long-term variation upon time. And, it enables to use the work spaces efficiently because many devices can be mounted vertically indoor of the dark box.

Unit Cell / Panel / Module

Multi-Channel I-V-L & Lifetime

Luminance / Color / Spectrum Measurement

Current / Power / Quantum Efficiency

Auto-Align & Auto-Focusing Control

Individual Temperature Control

Viewing Angle Measurement

2D Imaging Uniformity Measurement

High Space Utilization

Self-Developed Optical Detector

Manufacturer Calibration Solution


Division

Division

Product Name

M7000 PLUS Auto I-V-L Test System

Capacity

20ch / Unit Cell (Depends on Sample Size)

Motion Stage

Type

XYZ Auto Control Linear Motion (Option : Rotation Stage for Viewing Angle Measurement)

Stroke

X : 700mm, Y : 1,360mm, Z : 650mm

Resolution

±0.1mm

Vision

CCD Camera for Alignment and Focusing integrated in LUMO M20

Sourcemeter Unit

Model

Metro M131 LAB OLED Parameter Tester

Connection

Lifetime Tester & SMU

Voltage

Source Range

-20V ~ +20V

Source Resolution

5mV

Source Accuracy

(±0.05% of Set Value) ± 1mV

Measurement Range

-20V ~ +20V

Measurement Accuracy

(±0.05% of Measured Value) ± 1mV

Current

Configuration

5 Range Shunt Configuration (Auto Range)

Source Range

500mA

5mA

50μA

500μA

5nA

Source Resolution

500μA

5μA

50nA

500pA

5pA

Source Accuracy

(±0.05% of Set Value) ± (0.1% of Range Full Scale)

Measurement Range

500mA

5mA

50μA

500nA

5nA

Measurement Accuracy

(±0.05% of Set Value) ± (0.1% of Range Full Scale)

Motion Stage

Model

LUMO M20 Imaging Spectroradiometer

Wavelength Range

380~780 nm

Wavelength Precision

±0.3mm

Spectral Bandwidth

1.3nm or less

Measuring Angle

0.2° (Aperture 132mm)

Working Distance

132mm (Standard)

Luminance Range

0.1~100,000 cd/m2 (Incld. Auto ND Filter)

System Specification

System Configuration

  • Tech Icon
  • Leaflet
Tech Icon
Leaflet

M7000 PLUS Auto I-V-L Test System

Leaflet Download (2.69MB)

System Components

LUMO M20 Spectroradiometer

OLED Parameter Tester

Switching Unit

MC960 JIG

Auto I-V-L Test Software

DeJign Battery Test Jig

Q730 Battery Parameter Test Software

Q731 Battery Cycle Test Software