M131 LAB

OLED Parameter Tester


  • Component for OLED Electrical Test​
  • High Accurate, Stable Source and Measurement
  • ​4 Quadrant Operation Performance
  • ​Max. 30V Voltage Range / Max. 500mA Current Range / Max. 15Watt
  • ​Voltage Force Current Measurement (VFCM)
  • ​Current Force Voltage Measurement (CFVM)​
  • DC / AC Operation Mode
  • ​Fixed 1kHz Sine Wave for AC Operation 
  • Compliance Function for Protect Test Sample 
  • LAN Communication 

Technical Specification

Item

Specification

Model​

Metro M131 OLED Parameter Tester​

Control Mode​

VFCM, CFVM​

Connection

2 or 4 Probe

DC Characteristics (for IVL Measurement)​

Voltage​

Source Range​

Source Resolution​

Source Accuracy​

Measurement Range​

Measurement Accuracy​

-20V ~ +20V​

5mV​

(±0.05% of Set Value) ± 5mV

-20V ~ +20V​

(±0.05% of Measured Value) ± 5mV

Current​

Configuration​

Source Range​

Source Resolution​

Source Accuracy​

Source Accuracy​

Measurement Range​

Measurement Accuracy​

Measurement Accuracy​

5 Range Shunt Configuration (Auto Range)​

500mA​

10mA

100μA

1μA

10nA

500uA​

5uA​

50nA​

500pA​

5pA​

500mA ~ 10nA : (±0.05% of Set Value) ± (0.1% of Range Full Scale)

0 ~ 10nA : (±0.2% of Set Value) ± (0.1% of Range Full Scale)

500mA

10mA

100μA

1μA

10nA

500mA ~ 10nA : (±0.05% of Set Value) ± (0.1% of Range Full Scale)

0 ~ 10nA : (±0.2% of Set Value) ± (0.1% of Range Full Scale)

Measurement Interval​

Min. 100msec​

AC Characteristics (for CV Measurement)​

Operation Mode​

Sine Wave​

Frequency​

1kHz Fixed​

Sampling Rate​

100kSPS​

Data Point​

100 Point (Voltage, Current)​

Technical Specification